Quantcast
Channel: Data converters
Viewing all articles
Browse latest Browse all 38440

Forum Post: RE: Questions about TX-SDK-V2 EVB

$
0
0

Hi sir,

thanks for your reply .

sorry sir, you are absolutely right sir. Now with the Win xp I could get the GUI working.

Now to get start up,Individual Device Evaluation to test  LM96570 and LM96551. I followed Ultrasound Tx Evaluation Kit(TX_SDK_V1) User’s Guide 2.4 section.

1)Means power up sequence

2) launching app ,

3)connect, reset(enable and then disable), pll(enable and then disable),

4)after setting 2.4.5 point in user guide settings for beamformer controls,

5)connect, start Tx,

6)enabling high voltage pulser.

But I could not observe the output mentioned in 2.4.9 in oscilloscope at pulser o/p PS0-7 left side pin. I  Just observed noise at PS0-7 pins, and I tried to probe P0-7, N0-7 pins of beamformer which was same. Let me know sir, still any more information needed for my hardware set up.

could not getting where I am doing wrong. Please help me sir, it is very important for me.

And I observed the beamformer IC getting heated much than pulser, switch when I supplied power supplies especially 3.3v. Is it common or any problem is there?.


Thank you very much, waiting for your reply.


regards,

chandu.