Forum Post: RE: LMP91200MT pH Sensor

$
0
0

Hi Rahul,

I suggest the follwing approach:

Ph = 7 + Vout_LMP91200/alpha(T)
Vout_LMP91200 is measured in mV
alpha(T) = -(1/298)*T*59.16    (mV/Ph init)
T is measured in degKelvin

 When you read the output of the LMP91200, remeber to remove the VOCM contribution.

regards,

Domenico